Region

Applications

Glycols and Benzene in Toothpaste by FID on ZB-WAXplus

Application Detail (App ID: 16420)

Part No:7HG-G013-11
Phase:Polyethylene Glycol (PEG)
Dimensions:30 meters x 0.25 mm x 0.25 µm
Oven Profile:80 °C to 180 °C @ 20 °C/min for 5 min.
Carrier Gas:Constant Flow Helium, 3 mL/min
Injection:Split 20:1 0.5 µL @ 225°C
Detection: Flame Ionization (FID) (250°C)
Note:0.1 mg toothpaste in 1 ml 50:50 acetonitrile/H2O. 100 ul supernatant to 400 ul acetonitrile/water solvent. To 250 ul sample prep add 25 ul 5 mg/ml 1,3-propanediol (istd).

Analyte Details

Analyte Detail not available

1
Acetonitrile
search

2
1,2-Propanediol
search

3
Ethylene glycol (EG)
search

4
1,3-Propanediol
search

5
Glycerol
search