Component | As per USP | Adjustable from | Adjustable to | Limit |
---|---|---|---|---|
Mobile Phase Minor Component (<50%) | ±30% relative; cannot exceed ±10% absolute adjustment; Cannot be reduced to zero | |||
Mobile Phase pH | ±0.2 units | |||
Buffer Concentration (mM) | ±10% | |||
Column Temperature (°C) | ±10°C | |||
Flow Rate (mL/min) | ±50% (at same ID) |
As per USP | Desired column dimensions | |||
Column Length (mm) | ||||
Column Diameter (mm) | ||||
Particle Size (µm) | *Alternatively (as for the application of particle size adjustment to superficially porous particles), other L/dp combinations can be used provided that the number of theoretical plates (N) is within -25% to +50% | |||
L/Dp | % deviated | |||
The column with desired dimension | ||||
Relative Flow Rate Suggestion | ||||
Injection Volume (µL) | This is just a recommendation Can be adjusted as much as needed; must be consistent with linearity, precision, and detection requirements |
|||
Stationary Phase | No changes of the identity of the substituent recommended. |
Choose Phenomenex columns for your analysis |
||||
Choose USP column listing |