Region

Applications

Ganciclovir and Acyclovir on Luna 3µm HILIC with MS Detection

Application Detail (App ID: 16440)

Part No:00D-4449-B0
Dimensions:100 x 2 mm ID
Elution Type:Isocratic
Elution A:Acetonitrile
Elution B:100mM Ammonium Formate, pH 3.2
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 90 10

Flow Rate:0.4 mL/min
Col. Temp: ambient
Detection: Mass Spectrometer (MS) @ amu (22°C)
Detector Info: (MRM: Ganciclovir 256.2/152.3 blue, Acyclovir 226.2/152.2 red)
Note:The pH of the aqueous buffer is specified. The pH in mixed organic was not measured.

Analyte Details

Analyte Detail not available