SFC - Polar Selectivity Mix on Luna HILIC 250 x 4.6

Application Detail (App ID: 16489)

Part No:00G-4450-E0
Dimensions:250 x 4.6 mm ID
Elution Type:Gradient
Elution A:Methanol
Elution B:Carbon Dioxide
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 5 95
2 0.1 5 95
3 5 50 50
4 7 50 50

Flow Rate:3 mL/min
Col. Temp: 40 °C
Detection: UV-Vis Abs.-Fixed Wave. (UV) @ 254 nm (40°C)
Detector Info: Thar Discovery Series, Super Pure SFC instrument provided courtesy of Thar Technologies
Note:SecurityGuard™ Guard Cartridge System extends column lifetime.
- SecurityGuard Cartridges, HILIC 4 x 3.0 mm, 10/Pk Part No.: AJ0-8329
- Holder Part No.: KJ0-4282

Analyte Details

Analyte Detail not available

1
Benzoic acid
search

2
Caffeine
search

3
Acetaminophen
search