Region

Applications

Cocaine and Metabolites on Luna HILIC 3u 100x2.0

Application Detail (App ID: 17914)

Part No:00D-4449-B0
Dimensions:100 x 2 mm ID
Elution Type:Gradient
Elution A:95/5 ACN/100mM NH4 Formate pH 3.2
Elution B:50/45/5 ACN/H2O/100mM NH4 Formate pH 3.2
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 100 0
2 1.7 100 0
3 4.8 45 55
4 6.5 45 55
5 6.6 100 0
6 12.6 100 0

Flow Rate:0.4 mL/min
Col. Temp: ambient
Detection: Mass Spectrometer (MS) @ amu (22°C)
Detector Info: Mass Transitions
Cocaine 304 to 181.9
Norcocaine 290.2 to 168.2 & 290.1 to 136.0
Ecgonine methyl ester 200.2 to 182.1
Benzoylecgonine 290.2 to 168.2
Ecgonine 186.2 to 168.1
Note:SecurityGuard™ Guard Cartridge System extends column lifetime.
- SecurityGuard Cartridges, HILIC 4 x 2.0mm, 10/Pk Part No.: AJ0-8328
- Holder Part No.: KJ0-4282

Analyte Details

Analyte Detail not available

1
Cocaine
search

2
Norcocaine
search

3
Ecgonine methyl ester
search

4
Benzoylecgonine
search

5
Ecgonine
search