Region

Applications

Extraction of Cocaine-NIDA5 Using strata X-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS

Application Detail (App ID: 18843)

Part No:00B-4462-AN
Dimensions:50 x 2.1 mm ID
Elution Type:Gradient
Elution A:0.1% Formic acid in H2O
Elution B:0.1% Formic acid in ACN
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 90 10
2 2 30 70
3 2.01 5 95
4 2.3 5 95
5 2.31 90 10
6 4.1 90 10

Flow Rate:450 µL/min
Col. Temp: ambient
Detection: Mass Spectrometer (MS) @ amu (22°C)
Detector Info: ESI Condition
Positive Ionization
NEB: 12.00
CUR: 7.00
CAD: 12.00
IS: 5500.00
TEM: 400.00
EP 10.00

Analyte Details

Analyte Detail not available

1
Benzoylecgonine-d3
search

2
Benzoylecgonine
search

3
Cocaine
search

4
Cocaethylene
search