Region

Applications

USP Glycerin Purity Assay on ZB-BAC-2

Application Detail (App ID: 19291)

Part No:7HK-G022-32
Phase:Proprietary Blood Alcohol Phase
Dimensions:30 meters x 0.53 mm x 2 µm
Oven Profile:100 °C for 4 min to 120 °C @ 50 °C/min for 10 min to 220 °C @ 50 °C/min for 6 min
Carrier Gas:Constant Flow Helium, 4.5 mL/min
Injection:Split 10:1 1 µL @ 220°C
Detection: Flame Ionization (FID) (250°C)
Note:Concentrations were 2.0 mg/mL Glycerin, 0.050 mg/mL ethylene glycol, 0.050 mg/mL diethylene glycol, and 0.10 mg/mL 2,2,2-trichloroethanol (IS)
Run as a dual column analysis with ZB-BAC-1
Resolution = 11.09 between glycerin and diethylene glycol

Analyte Details

Analyte Detail not available

1
Ethylene glycol
search

2
2,2,2-Trichloroethanol
search

3
Diethylene glycol
search

4
Glycerin
search