Region

Applications

EtS EtG Standard

Application Detail (App ID: 19775)

Part No:00D-4387-Y0
Dimensions:100 x 3 mm ID
Elution Type:Gradient
Elution A:10mM Ammonium formate
Elution B:50:50 Methanol: Acetonitrile
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 100 0
2 4 20 80
3 4.1 100 0
4 7 100 0

Flow Rate:0.5 mL/min
Col. Temp: ambient
Detection: Mass Spectrometer (MS) @ amu (22°C)
Detector Info: SCIEX API 4000™

Analyte Details

Analyte Detail not available

1
EtG (Ethyl glucuronide)
search

2
Ethyl glucuronide-d5 (EtG-d5)
search

3
EtS (Ethyl sulfate)
search

4
EtS (Ethyl sulfate)
search