Region

Applications

Heterocyclic Amines by LCMS using Kinetex 5u C18 50x2.1mm

Sample Preparation Details

Part Number :8B-S029-UBJ
Description :Strata-X-C 33 µm Polymeric Strong Cation, 60 mg / 3 mL, Tubes, 50/Pk
Condition 1 :3 mL Methanol at 1 mL/min
Condition 2 :4 mL Water
Load :Load Sample
Wash 1 :3 mL 0.1M HCl
Wash 2 :3 mL Methanol
Elution 1 :3 mL 0.2M Ammonia Solution made up in methanol
Evaporate: :Evaporate to dryness
Reconstitute :0 µL Mobile phase

Application Detail (App ID: 20771)

Part No:00B-4601-AN
Dimensions:50 x 2.1 mm ID
Elution Type:Gradient
Elution A:30 mM Ammonium formate
Elution B:Acetonitrile
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 95 5
2 5 40 60
3 5.1 0 100
4 6 0 100
5 7 95 5

Flow Rate:500 µL/min
Col. Temp: ambient
Detection: Tandem Mass Spec (MS-MS) @ (22°C)
Detector Info: SCIEX API 4000™