Region

Applications

Quantification of Phenylurea Pesticides by QuEchERS and LC/MS/MS

Application Detail (App ID: 20894)

Part No:00B-4423-B0
Dimensions:50 x 2 mm ID
Elution Type:Gradient
Elution A:Water with 5 mM Ammonium Formate
Elution B:Methanol with 5 mM Ammonium Formate
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 80 20
2 8 10 90
3 14 10 90
4 15 80 20
5 20 80 20

Flow Rate:250 µL/min
Col. Temp: 25 °C
Detection: Tandem Mass Spec (MS-MS) @ (500°C)
Detector Info: 3200 QTRAP orAPI 3200

Analyte Details

Analyte Detail not available

1
Fenuron
search

2
Metoxuron
search

3
Thidiazuron
search

4
Tebuthiuron
search

5
Monolinuron
search

6
Chlortoluron
search

7
Isoproturon-d6
search

8
Fluometuron
search

9
Metabenzthiazuron
search

10
Metobromuron
search

11
Isoproturon
search

12
Diuron
search

13
Linuron
search

14
Propanil
search

15
Siduron
search

16
Diflubenzuron
search