Food - Quantitative Analysis of Quaternary Ammonium Compounds (QAC) by LC/MS/MS

Application Detail (App ID: 20913)

Part No:00F-4375-B0
Dimensions:150 x 2 mm ID
Elution Type:Gradient
Elution A:Water with 5mmol Ammonium Formate purified Water,
Elution B:Methanol with 5mmol Ammonium Formate
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 100 0
2 3 30 70
3 6 15 85
4 9 10 90
5 20.5 10 90
6 21 100 0
7 32 100 0

Flow Rate:100 µL/min
Col. Temp: 40 °C
Detection: Tandem Mass Spec (MS-MS) @ (425°C)
Detector Info: SCIEX API 4000™
Note:Courtesy of the laboratory CVUA Stuttgart (Schaflandstr. 3/2, 70736 Fellbach, Germany).