T3/T4 Plasma Extract by LC/MS/MS on Kinetex C18, 2.6 um, 75x2.1 mm

Sample Preparation Details

Part Number :8B-S029-UBJ
Description :Strata-X-C 33 µm Polymeric Strong Cation, 60 mg / 3 mL, Tubes, 50/Pk
Condition 1 :1 mL 100% MeOh at 1 mL/min
Condition 2 :1 mL 0.1 N HCl at 1 mL/min
Sample Pretreatment :Extraction Procedure: Activate with 1 mL MeOH Equilibrate the SPE cartridge with 1 mL 0.1 M HCl Load sample (0.4 mL serum/plasma + 1 mL 0.1 M HCl) Wash SPE with 1 mL 0.1 M HCl Wash SPE with 1 mL 100% MeOH Dry the SPE bed under maximum vacuum for 5 min Elute the analytes with 2x1 mL 2.5% NH4OH in MeOH Evaporate the solvent to dryness under gentle N2 stream at 40-45°C Reconstitute the dry residue in 200 uL initial MP (30%MeOH) solvent
Load :Load 0.10 mL Sample diluted in 0.1 1N HCl
Wash 1 :1 mL 0.1 N HCl at 1 mL/min
Wash 2 :1 mL 100% MeOH at 1 mL/min
Dry :5 min under max vac
Elution 1 :1 mL 2.5% NH4OH in MeOH at 1 mL/min
Elution 2 :1 mL 2.5% NH4OH in MeOH at 1 mL/min
Evaporate: :Evaporate to dryness
Reconstitute :200 µL 30% MeOH in 0.05% HCOOH

Application Detail (App ID: 20929)

Part No:00C-4462-AN
Dimensions:75 x 2.1 mm ID
Elution Type:Gradient
Elution A:0.05% Formic Acid in DI H2O
Elution B:50:50 MeOH:ACN
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 70 30
2 2.5 30 70
3 3 30 70
4 3.01 10 90
5 4 10 90
6 4.01 70 30
7 5.5 70 30

Flow Rate:600 mL/min
Col. Temp: ambient
Detection: Mass Spectrometer (MS) @ amu (0°C)
Detector Info: SCIEX API 5000™