Region

Applications

FAMEs In Walnut Oil by GC/MS on Zebron ZB-FAME

Application Detail (App ID: 24075)

Part No:7HG-G033-10
Phase:
Dimensions:30 meters x 0.25 mm x 0.2 µm
Oven Profile:100 °C for 2 min to 214 °C @ 3 °C/min
Carrier Gas:Constant Flow Helium, 1 mL/min
Injection:Split 50:1 1 µL @ 240°C
Detection: Mass Spectrometer (MS) (250°C)
Note:Recommended Accessories:
Recommended Liner: Zebron PLUS Single Taper Z-Liner, 4 mm ID
Liner Part No.: AG2-0A13-05 (for Agilent systems)
Inlet Seal: AG0-8620 (Gold Plated Easy Seal)
Septum: AG0-4696 (PhenoRed-400)

Sample Preparation:
1. Weigh out 25 mg of oil into a 5 mL screw cap glass vial
2. Add 2 mL of BF3-methanol to sample vial
3. Heat vial at 60 °C for 15 minutes with periodic vortexing
4. Remove from heat and allow to attain room temperature
5. Add 1 mL water followed by 1 mL hexane
6. Vortex for 1 minute
7. Remove upper organic layer
8. Dry hexane layer over anhydrous sodium sulfate
9. Transfer to autosampler vial and inject on GC

Analyte Details

Analyte Detail not available

1
C16:0
search

2
C18:1 cis 9
search

3
C18:2 cis 9,12
search

4
C18:3 cis 9,12,15
search