Region

Applications

IntantPC Labeled N-Linked Glycans from Infliximab-dyyb on bioZen Glycan

Application Detail (App ID: 25037)

Part No:00F-4773-AN
Dimensions:150 x 2.1 mm ID
Elution Type:Gradient
Elution A:100 mM Ammonium Formate, pH 4.5
Elution B:Acetonitrile
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 22 78
2 10 25.5 74.5
3 24 28 72
4 38.5 44.1 55.9
5 38.6 60 40
6 40.6 60 40
7 40.7 22 78
8 48 22 78

Flow Rate:0.5 mL/min
Col. Temp: 50 °C
Detection: LC/FLD (Fluorescence Detector) @ 285.0000000000 nm (nanometers) (0°C)
Note:Utilize bioZen N-Glycan Clean-Up microelution SPE 96well plate for sample preparation (Part No.: 8M-S009-NGA).

Analyte Details

Analyte Detail not available

1
G0-GlcNac
search

2
G1-GlcNac
search

3
G0
search

4
Man5
search

5
G0F
search

6
G1F-GlcNac
search

7
G1F
search

8
G1F'
search

9
G1FB
search

10
G2F
search

11
G2FB
search

12
Sialylated Glycan
search

13
Sialylated Glycan
search

14
Sialylated Glycan
search

15
Sialylated Glycan
search

16
Sialylated Glycan
search

17
Sialylated Glycan
search

18
Sialylated Glycan
search

19
Sialylated Glycan
search

20
Sialylated Glycan
search

21
Sialylated Glycan
search

22
Sialylated Glycan
search