Region

Applications

Identification and Quantification of Designer Drugs in Urine by LC-MS/MS

Application Detail (App ID: 25865)

Part No:00F-4462-Y0
Dimensions:150 x 3 mm ID
Elution Type:Gradient
Elution A:Ammonium formate in water
Elution B:Acetonitrile
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 95 5
2 6 5 95
3 6.2 5 95
4 8 95 5

Flow Rate:0.5 mL/min
Col. Temp: 50 °C
Detection: LC/MS/MS @ (0°C)
Note:To 20 µL urine, add 50 µL internal standard (D5-PCP in acetonitrile). Vortex then centrifuge for 10 min at 15,000 rpm.