Region

Applications

Phenylpropanolmine HCl-Limit of Amphetamine - Modified Method

Application Detail (App ID: 3275)

Part No:00D-4252-E0
Dimensions:100 x 4.6 mm ID
Elution Type:Isocratic
Elution A:Methanol
Elution B:THF
Elution C:2% tetramethylammonium hydroxide and 0.5% Phosphoric acid in water
Gradient Profile:
Step No. Time (min) Pct A Pct B Pct C
1 0 4 0.4 95.6

Flow Rate:1.5 mL/min
Col. Temp: ambient
Detection: UV-Vis Abs.-Variable Wave.(UV) @ 215 nm (22°C)
Note:SecurityGuard™ Guard Cartridge System extends column lifetime.
- SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287
- Holder Part No.: KJ0-4282

Analyte Details

Analyte Detail not available

1
Dextroamphetamine sulfate
search

2
Phenylpropanolamine
search