Region

How do I improve early eluting peak shape for GC?

TROUBLESHOOTING

ZEBRON GC

Last Updated: 7/29/2014

Many times early eluting compounds have poor peak shapes because the initial temperature may be too high and analytes are not initially concentrated (i.e. focused) at the front of the column. Polarity mismatch can also occur if a solvent doesn’t dissolve well in the phase. The solvent ‘pools’ on the column phase and is pushed down the column by the carrier gas. Some compounds travel with the solvent pools and may lead to tailing or split peaks.

 

Steps that can be taken to improve early eluting peak shape:

 

  • Use a split injection. This limits the amount of solvent that gets onto the column and reduces how much analyte dissolves in pooled solvent. The split injection also reduces peak widths because the higher flow through the inlet moves compounds from the inlet to the column faster. This will improve the peak shape considerably. The downside is that the split will sacrifice sensitivity. If you have sensitivity to spare or a more concentrated sample, this may be an option.
  • Decrease the injection volume. This also reduces the amount of solvent that is introduced onto the column. It also reduces the amount of analytes and therefore will also sacrifice sensitivity. Lower the initial oven temperature. This will force analytes to condense more into the phase and limits analyte migration during injection. This is known as analyte focusing. The downside of this is slightly longer cycle times because cooling to a lower temperature may take longer.
  • Use a pressure pulsed injection. This will increase the flow onto the column, but only temporarily. This helps focus the analytes and results in a narrower peak shape. Usually a pulsed injection will be about 10-15psi above the normal pressure for the duration of the injection. Afterward, the pressure and flow return to normal. Early eluting compounds can elute even earlier and a solvent delay may need to be reduced if using a MS detector.
  • Use a guard column. When used with volatiles, this is sometimes called a solvent gap. It can help to separate the analytes from the solvent and reduces the solvent pool effect. In some instances, it can also extend column lifetime because the guard can be trimmed or replaced when contaminated by dirty samples which would otherwise sacrifice the performance of the column.
  • Increase the column film thickness. A thicker film will better dissolve both volatile analytes and result in longer retention for early eluting peaks. A thicker film will also dissolve greater amounts of solvent and help minimize distortion due to solvent mismatch.