Abstract:
In this technical note, utilizing a Kinetex 2.7µm C18 100 x 2.1 mm column, we were able to provide acceptable, reproducible resolution for paracetamol and the principal degradation product impurity K by using the specific HPLC parameters (particle size, column dimension, and gradient mobile phase conditions) consistent with the Ph. Eur. impurities test for paracetamol published in Supplement 9.4.2-3 The Kinetex 2.7µm C18 column we used is a direct equivalent to the specified suitable column in Pharmeuropa 28.1 revision.
Document Type:
Technical Notes
Industries:
Pharmaceutical/Biopharmaceutical
Techniques:
HPLC / UHPLC
Brands:
Kinetex
PDF Link: