Region

Applications

EPA Method 537.1 - Gen X and Short Chain PFAS in Drinking Water

Application Detail (App ID: 25874)

Part No:00D-4752-AN
Dimensions:100 x 2.1 mm ID
Elution Type:Gradient
Elution A:0.1% Acetic Acid in Water
Elution B:Methanol
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 80 20
2 0.5 70 30
3 7 10 90
4 7.5 0 100
5 9 0 100

Flow Rate:0.7 mL/min
Col. Temp: 25 °C
Detection: LC/MS/MS @ (0°C)
Note:Drinking Water

Analyte Details

Analyte Detail not available