Component | As per Ph. Eur. monograph | Adjustable from | Adjustable to | Limit |
---|---|---|---|---|
Composition of the Mobile Phase + Gradient | Minor adjustments of the composition of the mobile phase and the gradient are acceptable, if the system suitability requirements are met, the principle peak(s) elute(s) within ± 15 % of the indicated retention time(s) and the final elution power of the mobile phase is not weaker. |
|||
Mobile Phase pH | No adjustment permitted |
|||
Buffer Concentration (mM) | No adjustment permitted |
|||
Column Temperature (°C) | ±5°C | |||
Flow Rate (mL/min) | Adjustment is permitted to maintain linear velocity when changing column dimensions |
|||
Column Length (mm) | ±70% | |||
Column Inner Diameter (mm) | ±25% | |||
Particle Size (µm) | No change permitted. |
|||
Stationary Phase | No change of the identity of the substituent permitted. |
Column dimension as per EP | Desired Column Dimensions | |||
Column Length (mm) | ||||
Column Inner Diameter (mm) | ||||
Flow Rate As per Ph. Eur. Monograph (mL/min) | ||||
Recommended Flow Rate To Maintain Linear Velocity (mL/min) |